KOMEG Technology Ind Co., Limited
|
Place of Origin: | Zhejiang, China (Mainland) |
Add to My Favorites | |
HiSupplier Escrow |
Quick Delivery Time Mini Style 64L Temperature Humidity Chamber
1. Standard configuration
a. Sight window*1, cable port (portΦ50mm)*1, shelves*2, chamber lamp*1, power cable *2m
b. No fuse breaker, over pressure,over heat and over current protection for compressor,
Over temperature protection, over load protect of blower, dry heat protection
2. Optional accessories
Cable Port: Φ100mm cable port is available
Shelves
Ro-type water purifier: R-80/day
Provide to keep the humidifying heater and wet bulb wick free from scale.
Inner glass door with operation port
A glass door can be provided behind the main door so that specimens can be observed. Two operation ports of 130mm diameter are used for handing specimens inside the chamber without opening the glass door.
Dehumidifier
The rotation regenerating dehumidifier M-120 ensures precise control of low humidity (5° C 5%RH) for electrostatic reliability tests.
Temperature and humidity recorder
-100 to +150° C/0 to 100%RH 100mm with effective width 100mm and 2 pen
Temperature 1 pen and humidity 1 pen.
-100 to +150° C/0 to 100%RH 100mm with effective width 100mm and 6 dots
Temperature 3 dots and humidity 1 dot.
-100 to +150° C/0 to 100%RH 100mm with effective width 180mm and 12 dots
Temperature 6 dots and humidity 6 dots.
Defrosting circuit
The chamber automatically detects and melts the frost on the evaporator when operating below 0° C.
Liquid or infection
In order to rapidly decrease the temperature inside the chamber, a cylinder of liquid can be connected the chamber.
3. Testing Graph (for example)
4. Quick Specification for Standard Climatic Test Chamber
Model | KMH-64L | |
64L chamber volume | ||
Temp. range | -40°C to +150°C | |
Display resolution | 0.1°C | |
Temperature stability | High ± 0.5°C ,low ± 1.0°C | |
Temperature uniformity | ±2.0°C | |
Heat up time | -40 °C to +100°C, within 45 min | |
Pull down time | +20 °C to -40°C, within 60 min | |
Humidity range | 20% to 98R.H, (standard), 10% to 98R.H, (with humidifier) | |
Humidity constancy | ±2.5R.H | |
Humidity uniformity | ±3.0R.H | |
Interior Dimension(mm) | 400W*400H*400D | |
Exterior Dimension(mm) | 600W*1526H*1200D | |
Interior material | Stainless Steel Plate ( SUS 304) | |
Exterior material | Baked Painting Steel Or Stainless steel ( SUS304 ) | |
Insulation material | Rigid polyurethane Foam and Glass Fiber Wool | |
Total power | 4.5KW ( heater 2KW included ) | |
Max current | 9.1A | |
Water Consumption | 0.35L / hr | |
Ambient Temp. | +5°C to +35°C | |
Power Supply | AC 380V±10 50Hz, 3 phase 4 wires +Ground Wires | |
Weight | 160KG | |
Sight window size | 215*230mm | |
Sample holder | 2 pcs | |
Fittings | cable port(portΦ50mm)*1, chamber lamp*1, power cable *2m | |
Safety device (standard) | No fuse breaker,over pressure,over heat&over current protect,over temp. protect, over load protect for blower, dry heat protect | |
Control System | Balanced Temperature & Humidity Control ( BTHC ) System | |
Refrigeration system | Air Cooled (Water Cooled is option) | |
Hermetic compressor | ||
Hermetic compressor Single stage refrigeration system, CFC-Free refrigerant (HFC-507 / HFC-23) | ||
CFC-Free refrigerant (HFC-507 / HFC-23) | ||
Controller | 7-inch LCD Touch Screen panel | |
Operation model: Program or fix point running | ||
Memory capacity: 120programs,1200steps, | ||
All repeat 999 cycles, part repeat 99 cycles | ||
Note: 1. We reserve the right to change specifications without prior notice 2. Customized sizes and configurations available |
5. Packing Info
We have standard inner box PP film and carton plywood for the packing. Especially for the samples,
we will pack them in much safe to guarantee that you can receive a please.
6. Installation Environment
Distance from the wall to both sides and rear of chamber more than 800mm, to the front more than 1500mm. Provide independent power distribution switchgear and humidification condensate drains, and external power connector device is necessary
Ground level, well-ventilated, non- flammable, explosive, corrosive gases and dust
No strong electromagnetic radiation nearby
With floor drain ( less than 2 meters from the refrigeration unit )
Venue floor load capacity: not less than 800kg per m2
Leave adequate space for maintenance