Post Offer Free
KOMEG Technology Ind Co., Limited  

Gold Index: 6311

You are here: home  > Customized 2000L Solar Panel Test Chamber for PV Module Dynamic Testing

Customized 2000L Solar Panel Test Chamber for PV Module Dynamic Testing 

Place of Origin: Zhejiang, China (Mainland) 
inquire
Add to My Favorites
HiSupplier Escrow
Share |

Product Detail

Model No.: KMH-2000

Customized 2000L Solar Test Chamber for PV Module Dynamic Testing
 
 
1. Application:<

Customized 2000L Solar Test Chamber for PV Module Dynamic Testing

 

 

1. Application:

 

Solar test chamber is mainly used for reliability test of solar cells in the specification to be verified in wet, frozen, temperature cycle, humidity test etc can be monitored during the experiment the continuity of the internal circuit of solar cells in order to determine whether the fault or failure, by the system solar battery failure can be determined. 

 

 

2. Features:

 

Data and Monitoring software

Module surface temp, test Chamber inside temp record and display

Module voltage & current record and display

With real time display curve and of the measured data and can be stored by USB connection

 

 

 

3. Test Model or test project

 

1). Damp Heat Test:

Test Condition:   95 ±2 ºC temperature range , and 95% ±5% R.H. humidity range

Test Duration: last for 1000H,

Test Purpose: To see the reliability of the module to withstand the effects of long-term penetration of heat and humidity.

 

2). Humidity-freeze test 

 

Test Purpose: to see the ability of the module to withstand the effects of high temperature and humidity followed by sub-zero temperatures.

Test points and requirements:

-40ºC lowest temperature and 85ºC highest temperature

Max temperature rising speed rate:200ºC/Hour=3.33/min

Max temperature falling speed rate:200ºC/Hour=3.33/min

24 hours per cycle, testing for 10 cycles, 240 hours (10 days)

Temp. Cycle Compare

High Temp. High Humidity

Low Temp

Temp. Change Rate

Cycle No.

IEC61215

85ºC/85±5%(20h)

-40ºC(0.5~4h)

High Temp. Rise& Fall(100ºC/h)
Low Temp. Rise& Fall(200ºC/h)

10

GB9535

85ºC/85±5%(20h)

-40ºC(0.5~4h)

High Temp. Rise& Fall(100ºC/h)
Low Temp. Rise& Fall(200ºC/h)

10

IEC61646

85ºC/85±5%(20h)

-40ºC(0.5~4h)

High Temp. Rise& Fall(100ºC/h)
Low Temp. Rise& Fall(200ºC/h)

10

GB18911

85ºC/85±5%(20h)

-40ºC(0.5~4h)

High Temp. Rise& Fall(100ºC/h)
Low Temp. Rise& Fall(200ºC/h)

10

UL1703

85ºC/85±2.5%(20h)

-40ºC(0.5~4h)

High Temp. Rise& Fall(100ºC/h)
Low Temp. Rise& Fall(200ºC/h)

10

IEC62108

85ºC/85±2.5%(20h)

-40ºC(0.5~4h)

High Temp. Rise& Fall(100ºC/h)
Low Temp. Rise& Fall(200ºC/h)

20
40

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3). Thermal cycle test;


Test requirements: -40ºC lowest temperature and 85ºC highest temperature
Temperature falling/rising speed rate:100ºC/Hour=1.67ºC/min,6 hours per cycle,

Test Duration: 50~200 cycles
total testing time last 300 to 1200 hours (12.5 to 50 days) 

 

Temp. Cycle Compare

Low Temp

High Temp

Temp Change RateºC/h

Dwell Time

Cycle Number.

IEC61215

-40±2ºC

85±2ºC

Max. 100ºC/h

Min.10min

50, 200

GB9535

-40±2ºC

85±2ºC

Max.100ºC/h

Min.10min

50, 200

IEC61646

-40±2ºC

85±2ºC

Max.100ºC/h

Min.10min

50, 200

GB18911

-40±2ºC

85±2ºC

Max.100ºC/h

Min.10min

50, 200

UL1703

-40±2ºC

90±2ºC

Max.120ºC/h

30~105min

200

IEC62108

-40±2ºC

65ºC,85ºC,110ºC

10cycle/day

10min

500,1000,2000

IEC62108

-40ºC

65ºC,85ºC,110ºC

18cycle/day

10min

500,1000,2000

IEEE1

bigPhoto
 Mr. Nancy Wong

Tel: 86-20-83971185
Contact to this supplier

Related Search

Find more related products in following catalogs on Hisupplier.com

Company Info

KOMEG Technology Ind Co., Limited [China (Mainland)]


Business Type:Manufacturer
City: Guangzhou
Province/State: Guangdong
Country/Region: China (Mainland)

You May Like:

Product (121)